Lightning, Connects directly to your Smartphone
This system measures the amount of blood glucose / β-ketone / total cholesterol (biochemical parameters) in whole blood. The test is based on the measurement of electrical current generated by the reaction of the biochemical parameters with the reagent of the strip. The meter measures the current, calculates the biochemical parameters in the blood, and displays the result on your mobile device. The strength of the current produced by the reaction depends on the amount of biochemical parameters in the blood sample.
Intended Use
This system is intended for use outside the body (in vitro diagnostic use) to quantitatively measure the biochemical parameters - blood glucose, β-ketone, total cholesterol - in whole blood. It is for self-testing. It should NOT be used for diagnosis or screening of diseases.
Price: AUD 99.95
LifeSmart MINI Multi Monitoring Device iOS | |
Model No. |
LS-946DI |
Dimension |
51.2 (L) x 37.68 (W) x 9.7 (H) mm |
Weight |
9.0 g |
Power Source |
iOS device |
Memory |
1 Set |
External Output |
Lightning connector Auto sample loading detection Auto electrode insertion detection Temperature Warning |
Operating Conditions |
10°C to 40°C (50°F to 104°F), between 10% and 85% R.H. (non-condensing) |
Meter Storage / Transportation Conditions |
-20°C to 60°C (-4°F to 140°F), between 10% and 93% R.H. (non-condensing) |
Strip Storage / Transportation Conditions |
2°C to 30°C (35.6°F to 86°F) and 10% to 85% R.H. (non-condensing) |
Measurement Units |
Blood Glucose / Total Cholesterol: mg/dL or mmol/L β-Ketone: Fixed mmol/L |
Measurement Range |
Blood Glucose: 10 to 700 mg/dL (0.56 to 38.89 mmol/L) β-ketone: 0.1 to 8.0 mmol/L Total cholesterol: 100 to 400 mg/dL (2.5 to 10.3 mmol/L) |
Expected service life |
5 years |
Operating Altitude |
Up to 2000m, for indoor use |
Degree of Pollution |
Pollution degree 2 |
Strip I - Cholesterol Strip | |
Enzyme Type |
Cholesterol esterase (Pseudomonas sp.) ≥ 0.1 U; Cholesterol oxidase (Streptomuces) ≥ 0.1 U |
Sample Size |
3 μL |
Reaction Time |
≤ 60 seconds |
Parameter |
Cholesterol |
Measurement Range |
100 to 400 mg/dL (2.6 to 10.4 mmol/L) |
Hematocrit Range |
20% - 60% |
Precision |
CV < 7.5% |
Package |
Single Foil Pack |
Strip II - Glucose and Hematocrit Strip | |
Enzyme Type |
GDH-FAD |
Sample Size |
1.0 μL |
Reaction Time |
5 seconds |
Parameter |
Glucose |
Measurement Range |
10 - 700 mg/dL (0.56 - 38.89 mmol/L) |
Hematocrit Range |
0% - 70% |
Precision |
∓ 5% with respect to standard |
Accuracy |
∓ 15 mg/dL if ≦ 100 mg/dL; ∓ 15 % if > 100 mg/dL |
Package |
Vial Pack |
Strip III - Ketone Strip | |
Enzyme Type |
HBD |
Sample Size |
1.0 μL |
Reaction Time |
10 seconds |
Measurement Range |
0.1 - 8.0 mmol/L (1.04 - 83.2 mg/dL) |
Hematocrit Range |
10% - 70% |
Precision |
≦ 1 mmol/L, SD < 0.1 mM; > 1 mmol/L, CV < 7.5% |
Package |
Single foil pack |